|
Using ACR 'Suspicious' Subcategories of 4(a), 4(b) and
4(c)
Since
the ACR introduced the new 'Suspicious' subcategories of 4(a), (b) and
(c), some facilities have chosen to replace the 'Suspicious' category
with one of the new subcategories. This is not allowed under MQSA
regulations. Both the FDA and the ACR have addressed the
situation.
From the FDA website Policy Guidance Help System:
Question 18: The American College of Radiology Breast
Imaging Reporting and Data System (BIRADS) suggests that facilities
subdivide the “Suspicious” assessment category into one of three
subcategories (4A-Low Suspicion for Malignancy, 4B-Intermediate
Suspicion for Malignancy, and 4C-Moderate Concern but not Classic for
Malignancy). Can facility reports use these subcategories instead of the
“Suspicious” assessment category?
No. While the facility has the option of using one of
the three subcategories in addition to a final assessment of
“Suspicious”, it cannot use the subcategories instead of the
“Suspicious” assessment category on the mammography report
Following are excerpts from the ACR BIRADS frequently
asked questions posted on the ACR website:
|
BIRADS Category |
Assessment |
Followup Recommendations |
|
4
|
Suspicious Abnormality - Biopsy Should Be Considered |
Usually requires biopsy |
| Optional subdivisions:* 4A: Finding
needing intervention with a low suspicion for malignancy
4B: Lesions with an intermediate
suspicion of malignancy
4C: Findings of moderate concern, but
not classic for malignancy |
*A subdivision may be used in
addition to the Category 4 final assessment; MQSA does
not allow a subdivision to replace a Category 4 final
assessment. Use of subdivision is at the discretion of the
facility it is not required by the FDA.
Q. We wish to use the ACR's Category 4 subdivisions (4A - Finding
needing intervention with a low suspicion for malignancy, 4B - Lesions
with an intermediate suspicion of malignancy, and 4C - Findings of
moderate concern, but not classic for malignancy). Can our reports
use these subcategories instead of the Category 4 assessment (
Suspicious Abnormality - Biopsy Should Be Considered)?
A. No. While you have the option of using one of the 3
subcategories in addition to a final assessment of "Suspicious",
the FDA will not allow you to use the subcategories instead of
the "Suspicious" assessment category on the mammography report.
(This question and answer was adapted from the FDA's
Policy Guidance Help System.)
link to ACR BIRADS FAQ
* * * * * * * * * * * * * * * *
Continuing Experience Documentation for Interpreting
Physicians Reading at Multiple
Sites
Comments:
Physicians continuing experience
documentation may include the following:
§ A letter, table, or
printout from each facility
§ Signed by a responsible
facility official
§ Stating that he/she has
interpreted a given number of mammograms at the facility in a given time
period
The inspector will no longer accept
printouts from the interpreting physician’s office that include all
sites where they read.
*The only exception to this would be
if the ownership of the facilities where the interpreting physicians
read is the same person(s) signed by the same responsible facility
official.
For example: If a radiologist’s office
owns 5 different sites where they read and one coordinator who keeps up
with their continuing experience for all sites, that coordinator can
provide a list documenting their continuing experience. This list must
be signed by a responsible facility official who may be the same person
if the official is employed by the owner of all of the facilities.
* * * * * * * * * * * * * * * *
When a Facility Has More Than One
Processor
The annual survey or equipment evaluation
requirement for all clinically used mammography processors is addressed
in MQSA as the following:
21 CFR 900.12 (e)(9)(i) At least once a
year, each facility shall undergo a survey by a medical physicist or by
an individual under the direct supervision of a medical physicist. At a
minimum this survey shall include the performance of tests to ensure
that the facility meets the quality assurance requirements of the annual
test described in 900.12(e)(5) and (6) and the weekly phantom image
quality test described in 900.12(e)(2).
21 CFR 900.12(e)(10) Mammography
equipment evaluations. Additional evaluations of mammography units or
image processors shall be conducted whenever a new unit or processor is
installed, a unit or processor is dissembled and reassembled at a new
location or major components of a mammography unit or processor
equipment are changed or repaired. These evaluations shall be used to
determine whether the new or changed equipment meets the requirements of
applicable standards in 900.12(b) and (e). All problems shall be
corrected before the new or changed equipment is put into service for
examinations or film processing. The mammography equipment evaluation
shall be performed by a medical physicist or by an individual under the
direct supervision of a medical physicist.
Comments:
The annual survey or
equipment evaluation must include all processors used clinically, even
those at remote sites or back-up processors. It is the
responsibility of the facility to inform the medical physicist prior to
introducing a different processor into mammography and at the annual
survey. The equipment evaluation requirement for the processors is
the following:
§
Sensitometric testing as described
in 900.12(e)(1)
§
Phantom testing as described in
900.12(e)(2)
§
System artifact evaluation as
described in 900.12(e)(5)(ix)
§
Dose determination as described in
900.12(e)(5(vi) if clinical techniques increase significantly
§
Verification of the
appropriate processing solutions as described in 900.12(b)(13)
****PLEASE
BE CERTAIN THAT YOU ARE PERFORMING A WEEKLY PHANTOM ON YOUR SECOND
PROCESSOR IF BOTH PROCESSORS ARE USED REGULARLY FOR MAMMOGRAPHY****
* * * * * * * * *
* * * * * * *
.
Uniformity of Screen Speed
The uniformity
of screen speed requirement is addressed in MQSA as the following:
21 CFR 900.12 (e)(5)(viii): Uniformity of screen speed of all the
cassettes in the facility shall be tested and the difference between the
maximum and the minimum optical densities shall not exceed 0.30. Screen
artifacts shall also be evaluated during this test.
Comments:
Please be aware that if the physicist does not document for screen
artifacts during this test, the facility will be in violation and cited.
Also, Uniformity of Screen Speed MUST be performed on all
NEW screens before they are put into service. Failure to do so
will result in a violation. This may be performed by the physicist
on site, or by physicist oversight.
* * * * * * * * * * * * * * * *
Alternative
Standards for Processor QC
Conducting the
daily processor QC when the sensitometer is not available.
On 10-18-1999,
FDA approved a request for an alternative to sensitometric-densitometric
testing of processor performance that can be used for a period of up to
2 weeks when the facility’s sensitometer is unavailable.
When using the alternative test, processor performance is considered
satisfactory if:
1. The optical
density of the film at the center of an image of a standard FDA-accepted
phantom is at least 1.20 when exposed under typical clinical conditions.
2. The optical
density of the film at the center of the phantom image changes no more
than +/- 0.20 from the established operating level.
3. The density
difference between the background of the phantom and an added test
object, used to assess image contrast, is measured and does not vary by
more than +/- 0.05 from the established operating level.
********IN
ADDITION********
4. To evaluate
base+fog an additional measurement of density must be made either of a
shielded portion of the phantom image film or of an unexposed film. In
accordance with 21 CFR 900.12 (e)(1)(i), the base+fog density must be
within +0.03 of the established operating level.
As with the
original test, this alternative test must be conducted “each day
clinical films are processed, but before processing of clinical films.”
All results must be recorded. Again, as with the original test, if
processor performance fails to meet any part of the alternative test,
the problem must be corrected before processing is resumed.
|