| This alternative standard was approved on May 31, 2002. It
defines the conditions under which the mammography equipment
evaluations performed after some
computer software upgrades may be performed either by a
medical physicist on site or under the conditions of Medical
Physicist Oversight. If these conditions are not met the mammography
equipment evaluation after the upgrade must be performed by a
medical physicist on site. The original standard is contained
within 21 CFR 900.12(e)(10).
(10) Mammography equipment
evaluations. Additional evaluations of mammography units or image
processors shall be conducted whenever a unit or processor is
installed, a unit or processor is dissembled and reassembled at the
same or a new location, or major components of a mammography unit or
processor equipment are changed or repaired. These evaluations shall
be used to determine whether the new or changed equipment meets the
requirements of applicable standards in paragraphs (b) and (e) of
this section. All problems shall be corrected before the new or
changed equipment is put into service for examinations or film
processing. The mammography equipment evaluation shall be performed
by a medical physicist or by an individual under the direct
supervision of a medical physicist.
The approved alternative and the conditions for its use are:
Software changes or upgrades are considered by FDA to be major
repairs, thus the facility must have a mammography equipment
evaluation performed after installation of such a change or upgrade.
The mammography equipment evaluation must be performed and all
failures to meet the applicable standards must be corrected before
the affected equipment is used for patient examinations. The tests
to be included in the mammography equipment evaluation must be
specified by the manufacturer. The specified tests must be adequate
for determining whether all of the standards of 21 CFR 900.12(b) and
(e) that are applicable to the upgrade are met. If the tests
included in the mammography equipment evaluation are all tests that
are performed by the quality control technologist as part of the
quality assurance program required by the manufacturer, then the
mammography equipment evaluation may be conducted either during an
onsite visit by a medical physicist or under Medical Physicist
Oversight. If any of the necessary tests after the software upgrade
are required to be performed by the medical physicist, the
mammography equipment evaluation must be performed in its entirety
by the medical physicist on site.
Additional conditions for using this alternative requirement in
association with a software upgrade are that:
-
The manufacturer must notify FDA of its
intention to install the upgrade. The notification must include
a brief description of the upgrade, the model(s) of the units
that will be upgraded, and a copy of the information to be
provided to each facility describing the upgrade and the
facility’s post installation responsibilities. The manufacturer
must receive confirmation from FDA that the upgrade is covered
by the alternative requirement before beginning installation.
-
By the completion of each individual upgrade,
the manufacturer must inform the facility in writing of its post
installation responsibilities under the alternative requirement,
which are that the facility must:
-
conduct a mammography equipment evaluation after
installation of the upgrade, either during a medical physicist
onsite visit or under Medical Physicist Oversight,
-
include in its mammography equipment evaluation
the tests specified by the manufacturer,
-
perform the mammography equipment evaluation and
correct all test failures before the affected equipment is used
for patient examinations, and
-
keep records of the test results and follow-up
actions in accordance with 21 CFR 900.12(d)(2).
|