| This alternative standard was approved on December 14,
2004. It covers the system artifact testing at remote sites where
mobile mammography facilities utilize processors that are
permanently located at those sites. The approved alternative permits
a specially trained quality control technologist to make system
artifact films and phantom images at remote processor sites used by
mobile mammography facilities and submit them to the facility
medical physicist for evaluation. This relieves the facility from
the need to have the medical physicist visit each remote processing
location as a part of the annual survey.
Although this request was approved for a
specific mobile facilities operator, it has been extended to cover
any other mobile facility that elects to apply it and meets the
conditions described below. The alternative was approved for an
indefinite period.
The original standard states:
21 CFR 900.12(e)(5)(ix): System Artifacts.
System artifacts shall be evaluated with a high-grade,
defect-free sheet of homogeneous material large enough to cover
the mammography cassette and shall be performed for all cassette
sizes used in the facility using a grid appropriate for the
cassette size being tested. System artifacts shall also be
evaluated for all available focal spot sizes and target filter
combinations used clinically.
The above is required to be performed annually by a qualified
mammography medical physicist under the following:
21 CFR 900.12(e)(9)(i): At least once a year each facility
shall undergo a survey by a medical physicist or an individual
under the direct supervision of a medical physicist. At a
minimum, the survey shall include the performance of tests to
ensure that the facility meets the quality assurance
requirements of the annual tests described in paragraphs (e)(5)
and (e)(6) of this section and the weekly phantom image quality
test described in paragraph (e)(2) of this section.
The granted alternative is:
21 CFR 900.12(e)(9)(i): At least once a year each facility
shall undergo a survey by a medical physicist or an individual
under the direct supervision of a medical physicist. At a
minimum, the survey shall include the performance of tests to
ensure that the facility meets the quality assurance
requirements of the annual tests described in paragraphs (e)(5)
and (e)(6) of this section and the weekly phantom image quality
test described in paragraph (e)(2) of this section. For mobile
facilities using remote site processing, a mammography
technologist at the facility will be trained by the medical
physicist to produce the processor artifact test images at each
remote site and will provide them to the medical physicist for
evaluation during the annual survey.
Conditions of this alternative requirement are:
The medical physicist will train the mammography technologist in
the procedures as outlined in the American College of Radiology (ACR)
Mammography Quality Control Manual, 1999. The training will include
the direct supervision of that mammography technologist while she
carries out these procedures at a mobile mammography site. The
medical physicist will also carry out the identical procedures at
that site. These procedures are:
- Expose and process two mammography films according to the
artifact test procedure described in the 1999 ACR Manual in
accordance with the medical physicist’s instructions.
- Expose and process a phantom image.
The medical physicist will evaluate and compare all these test
films produced by the mammography technologist and by the medical
physicist. If the results are comparable and the medical physicist
reaches the same conclusions after reading the films produced by the
mammography technologist that would be reached reading the
physicist’s films, then the mammography technologist is considered
to be capable of producing adequate films for this test without
direct supervision.
At least once a year, the physicist will perform all required
annual survey items including the phantom image and artifact test at
one location serviced by the medical physicist and will use the
technologist’s ACR phantom and absorber materials to perform the
on-site testing.Prior to this annual visit, the mammography
technologist will take two artifact films using four cm of absorber
provided by the physicist at each remote processing site. The films
will be developed according to the procedures in the 1999 ACR
manual. The artifact test procedures will be repeated for each
target/filter combination and each processor used clinically at each
site. The mammography technologist will also take a phantom image at
each site using the facility’s ACR phantom. The images taken by the
mammography technologist at the remote sites will be brought to the
chosen annual survey site for evaluation by the medical physicist
during the annual survey. Applicable records for the physicist’s
“Evaluation of Site’s Technologist’s QC Program” as described in the
1999 ACR Mammography Quality Control Manual, will also be provided
from each remote site for the physicist’s review.
|