Reducing
Radiation Exposure to the Citizens of North Carolina |
![]() |
|
N. C. Regulations for Protection Against Radiation |
|
|
|||||
|
|
|||||
|
15A NCAC 11 .0359 MEASUREMENTS/DOSAGES OF UNSEALED RADIOACTIVE MATERIAL FOR MEDICAL USE (a) A licensee shall possess and use a dose calibrator to measure the radioactivity of dosages of photon-emitting radionuclides prior to administration to each individual. A licensee shall: (1) develop, maintain, and implement written procedures for use of the dose calibrator; (2) calibrate each dose calibrator in accordance with the requirements of 10 CFR 35.60(b). (b) A licensee shall retain a record of each check, test, and calibration performed in accordance with this Rule for a period of three years following the test.
History Note: Authority G.S. 104E-7; 104E-10(b); 104E-12; Eff. April 1, 1999; Amended Eff. November 1, 2007.
|
|||||
|
|
|
|
|
||
|
|
|||||